This hole serves two purposes – to allow replacement of the electrolyte and also to allow pressure to be equalised during measurement. If the hole is covered during measurement then the electrolyte will not be able to travel through the electrode’s diaphragm into the sample thus the electrode accuracy and responsiveness will be reduced.