For an ideal electrode the potential difference between the reference and measuring electrode is 0mV at pH 7. As the ideal electrode doesn’t exist the deviation of a few mV at pH 7 that is found with real electrodes is called the asymmetry potential, Eo. In practical terms it is a measure of the performance of the reference electrode – if the result is outside of +/- 15mV this is indicative of problems with the reference electrode such as contaminated electrolyte or reference element poisoning.